Measuring Ethylene on Perkin-Elmer GC |
||
11-3-97 TSA |
||
1. Detector A (top) is for whole cell assays and detector B is for enzyme assays. | ||
2. Turn D to "column heater", left light is on. Push start ---> B light on. | ||
3. Set "detector air" at 20 PSI. | ||
4. "Detector A hydrogen" up, adjust hydrogen to 20 PSI. | ||
5. "Detector A hydrogen" down, "detector polarity" at A+, push "ignite" and lift "detector A hydrogen" at the same time. Pop sound signifies that the GC has ignited. | ||
6. G at 1024X, H at X1K. Warm up for 20 minutes. | ||
7. Zero chart recorder, put G and H to 1X, then use E and F to zero. | ||
8. Flush needle with N2, inject with 0.1 ml of sample. | ||
9. CH4 comes out with a shadow first, then ethylene (C2H4). Acetylene (C2H6) has a huge peak about 20 minutes later; therefore, inject as many samples as possible within that time. | ||
10. Turn off: | ||
"detector A hydrogen" | down | |
hydrogen | off | |
air | off | |
D | off | |
chart recorder | off | |